Title of article :
MR relaxometry on a whole-body imager: quality control
Author/Authors :
H?jek، نويسنده , , Milan and Babi?، نويسنده , , Milan and Herynek، نويسنده , , V??t، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
1087
To page :
1092
Abstract :
Long-term monitoring of the average proton relaxation time T2 of phantoms measured on a Siemens MR whole-body imager showed very good repeatability and reproducibility. The repeatability (short-term precision) and reproducibility (long-term precision) of the average values of a relaxation time T2 ≅ 81 ms, obtained by a standard 16-echo CPMG pulse sequence, were 2.6% and 9.7%, respectively. The Siemens Vision imager proved to be a suitable tool for T2 evaluation in vivo. Quality control was performed using the techniques of control diagrams developed by Shewhart, which proved to be an appropriate method for continuous quality control of relaxation time determination.
Keywords :
t2 , Short- and Long-term precision , Shewhart’s control charts