• Title of article

    Phantom and in vivo study of the look–locher T1 mapping method

  • Author/Authors

    Karlsson، نويسنده , , Magnus and Nordell، نويسنده , , Bo، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    8
  • From page
    1481
  • To page
    1488
  • Abstract
    This paper describes and tests the LL-EPI method for obtaining quantitative T1 estimates in a few seconds thereby allowing dynamic T1 studies. It is shown that the method works even when there is an inflow into the imaged volume, e.g., in a vessel. No calibration is needed. The method has been tested in a phantom study with several different scan parameter set-ups, with and without inflow. The method shows robustness and individual scan parameters and inflow rates do not influence the ability to calculate the Gd-DTPA concentration. Linearity prevail between the measured 1/T1 and the Gd-DTPA concentration in the range 150 < T1 < 2500 ms. In a dynamic Gd-DTPA phantom study, it was shown that the dynamic LL-EPI T1 mapping technique was three times more sensitive than the signal from a T∗2-weighted EPI sequence. In an in vivo study, dynamic T1 mapping of the Gd-DTPA uptake in a meningioma was performed. Inspection of the uptake curves indicates that the method is feasible in clinical perfusion studies.
  • Keywords
    Dynamic studies , Gd-DTPA , Relaxivity , T1 mapping
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1999
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1830423