Title of article :
MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging
Author/Authors :
Hedlund، نويسنده , , Laurence W and Cofer، نويسنده , , Gary P and Owen، نويسنده , , Steven J and Allan Johnson، نويسنده , , G، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
We describe an MR-compatible ventilator that is computer controlled to generate a variety of breathing patterns, to minimize image degrading effects of breathing motion, and to support delivery of gas anesthesia and experimental inhalational gases. A key feature of this ventilator is the breathing valve that attaches directly to the endotracheal tube to reduce dead volume and allows independent control of inspiratory and expiratory phases of ventilation. This ventilator has been used in a wide variety of MR and x-ray microscopy studies of small animals, especially for MR imaging the lungs with hyperpolarized gases (3He & 129Xe).
Keywords :
MR microscopy , respirator , Mechanical Ventilation , Rodents , Xenon , Helium , Hyperpolarized gas
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging