Title of article :
A high-resolution phantom for MRI
Author/Authors :
Fellner، نويسنده , , Claudia and Müller، نويسنده , , Walter and Georgi، نويسنده , , Jens and Taubenreuther، نويسنده , , Ulrike and Fellner، نويسنده , , Franz A and Kalender، نويسنده , , Willi A، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
Assessment of spatial resolution is an important step to test the performance of new sequence techniques-especially ultrafast techniques with dedicated k-space trajectories or interpolation algorithms. Measurement of the modulation transfer function (MTF) is a rather difficult procedure, but using suitable resolution phantoms allows a simple visual evaluation of spatial resolution.
trast to commonly used test objects with a very restricted number of resolution patterns we developed a phantom containing resolution patterns from 0.1 to 1.5 mm in steps of 0.1 mm. One resolution pattern consists of five parallel Plexiglas strips with the distance of the strips being equal to their thickness. Together with a Plexiglas cuboid the resolution patterns are mounted on a Plexiglas plate on the bottom of the cylindrical phantom. An aqueous solution of manganese chloride is used to fill the phantom. High resolution cross sections (pixel size: 50 μm) through the resolution patterns were measured to confirm the correct dimensions of the phantom. To verify the appropriateness of the 0.1 and 0.2 mm stacks micro-CT images with a pixel size of 25 μm were acquired additionally for both patterns. Besides visual inspection evaluation of the profile function of signal intensity across the stacks demonstrates that the resolution patterns are sufficiently correct. T1-weighted SE sequences with slightly different pixel sizes as well as T1- and T2∗- weighted gradient echo sequences were applied to demonstrate some possible applications of this phantom. In conclusion, the proposed phantom is well suited to assess the spatial resolution qualitatively (i.e., visually) and quantitatively over a wide range in steps of 0.1 mm.
Keywords :
MRI , Phantom , Spatial resolution , high resolution , quality assurance
Journal title :
Magnetic Resonance Imaging