Title of article
Low field thoracic MRI—a fast and radiation free routine imaging modality in children
Author/Authors
Wagner، نويسنده , , M. and Bِwing، نويسنده , , B. and Kuth، نويسنده , , R. and Deimling، نويسنده , , M. and Rascher، نويسنده , , W. and Rupprecht، نويسنده , , T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
9
From page
975
To page
983
Abstract
Radiography of the chest is the most frequently performed radiological examination in pediatric imaging. However, it is associated with the application of ionizing radiation. In order to avoid ionizing radiation in children a new and very fast MRI technique has been developed at our center as an alternative to the pediatric chest X-ray. 100 patients who had received a chest X-ray were additionally investigated in a 0.2 T low-field MR-scanner by a modified true FISP sequence with an acquisition time of 3.6–4.6 s for a coronal triple-slice scan. X-ray and MR images were independently evaluated and later compared by two pediatric radiologists. Total investigation times (door-to-door time) for X-ray and MRI were comparable. The signal-to-noise ratio for lung parenchyma was 4.6–7.3. Of 189 pathologic findings 165 were depicted on MR images as well as radiographs, 18 were noted on MRIs only, 6 on X-rays only. Overall kappa (κ) was 0.87. True FISP MRI may be a good alternative to conventional chest X-ray. The main advantages are: fast imaging free of ionizing radiation, easy performance, no need for special equipment, optional imaging in all 3 planes, good image quality, and a high diagnostic value.
Keywords
MAGNETIC RESONANCE IMAGING , Lung , Diagnostic Imaging , children , Diagnostic techniques , Lung disease , Respiratory System
Journal title
Magnetic Resonance Imaging
Serial Year
2001
Journal title
Magnetic Resonance Imaging
Record number
1831135