Title of article :
A sixteen-channel multiplexing upgrade for single channel receivers
Author/Authors :
Porter، نويسنده , , Jay R. and Wright، نويسنده , , Steven M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
With the increasing interest in phased arrays in magnetic resonance imaging, imaging system receivers capable of acquiring larger number of parallel signals are needed. Suggested techniques for rapid imaging propose the use of arrays with as many as 128 elements. While simply duplicating the number of receiver chains as needed is a viable technique, it quickly becomes both cumbersome and expensive.
omain multiplexing offers an alternative solution to this problem. By using RF multiplexing switches, a single receiver can be upgraded to an array receiver capable of multi-channel data acquisition giving users array capability. Additionally, it can be used to dramatically increase acquisition capability of multiple receiver systems.
aper reports results from a multiplexing system upgrade, which converts a single channel standard clinical imaging system to a 16-channel array system. The upgrade includes both the RF multiplexing front-end and an external data acquisition system with image processing capability. Issues concerning the implementation of high channel-count multiplexers are also discussed.
Keywords :
phased arrays , multiplexer , MAGNETIC RESONANCE IMAGING
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging