• Title of article

    Real time MRI-ultrasound image guided stereotactic prostate biopsy

  • Author/Authors

    Kaplan، نويسنده , , Irving and Oldenburg، نويسنده , , Nicklas E. and Meskell، نويسنده , , Paul C Blake، نويسنده , , Michael W. Church، نويسنده , , Paul and Holupka، نويسنده , , Edward J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    5
  • From page
    295
  • To page
    299
  • Abstract
    To report a technique for target directed transperineal ultrasound guided biopsy using high resolution endorectal MRI images Ultrasound fusion. Two patients presented after external beam irradiation for prostate cancer with a rising PSA. An Endorectal MRI using a 1.5 Tesla scanner was obtained. Subsequently a Transrectal Ultrasound guided biopsy was performed. The Ultrasound probe was fixed to a stepper-stabilizer to provide a reference coordinate system for stereotaxic needle biopsy needle placement. The MRI image set was fused to the Ultrasound images in real time. Abnormal areas determined in the MR images were targeted for biopsy. Recurrent prostate carcinoma was detected pathologically in 3 of 4 stereotactic biopsies. Abnormal areas suspicious for cancer detected on T1 weighted images obtained in a strong field Endorectal MRI scan can be targeted for stereotactic biopsy using Transrectal Ultrasound. This image guide technique may be very useful in directing biopsies.
  • Keywords
    prostate cancer , MAGNETIC RESONANCE IMAGING , biopsy , Ultrasound
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2002
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1831370