Title of article :
Multicenter trial for the set-up of a MRI quality assurance programme
Author/Authors :
Colombo، نويسنده , , P. and Baldassarri، نويسنده , , A. and Del Corona، نويسنده , , M. and Mascaro، نويسنده , , L. and Strocchi، نويسنده , , S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
9
From page :
93
To page :
101
Abstract :
Many international protocols related to RMI-QC program are focused on acquisition methods and analysis of several image quality parameters but rarely normality ranges or measurement frequencies are presented. To address this problem we investigated the variability of many magnetic resonance imaging (MRI) systems with the set-up of multicenter trial. The trial was set up to investigate short-and mid-term variability of two fundamental nongeometric image quality parameters: signal-to-noise (SNR) and integral percent uniformity (U%). Ten centers (12 devices) participated to data collection consisting of a three-step-protocol. First, 10 consecutive images of a phantom were collected with a spin echo sequence. As second step the series collection was repeated 24 h later. Finally a single image acquisition was performed twice a week for 5 weeks. The analysis of results allowed us to define a “physiological” variability of ±3% of the reference level for both parameters and to conclude that a weekly measurement is adequate to detect relevant variations of device performance.
Keywords :
quality assurance , Uniformity , Signal-to-noise ratio
Journal title :
Magnetic Resonance Imaging
Serial Year :
2004
Journal title :
Magnetic Resonance Imaging
Record number :
1831833
Link To Document :
بازگشت