Title of article :
Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T
Author/Authors :
Shellock، نويسنده , , Frank G. and Valencerina، نويسنده , , Samuel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
5
From page :
1021
To page :
1025
Abstract :
Specialized implants are used for transcatheter closure of septal defects, including atrial and ventricular septal defects, and patent foramen ovale. These metallic devices may pose a risk to patients undergoing magnetic resonance imaging (MRI) procedures especially if performed at 3 T. Therefore, this investigation evaluated MRI safety at 3 T for septal repair implants (CardioSEAL Septal Repair Implant and STARFlex Septal Repair Implant, NMT Medical, Boston, MA, USA) by characterizing magnetic field interactions, heating and artifacts. These implants exhibited minor magnetic field interactions; heating was not excessive (+0.5°C); and artifacts will only create a problem if the area of interest is in the same area as or near these devices. Thus, the findings indicated that it would be safe for a patient with these implants to undergo MRI at 3 T or lower. Importantly, because of the minor magnetic field interactions, MRI may be performed immediately after implantation.
Keywords :
Ventricular septal defect , Artifacts , Atrial Septal Defect , Implants , Safety , Patent foramen ovale , Septal repair , Magnetic Resonance Imaging , specific absorption rate
Journal title :
Magnetic Resonance Imaging
Serial Year :
2005
Journal title :
Magnetic Resonance Imaging
Record number :
1832206
Link To Document :