• Title of article

    HASTE sequence with parallel acquisition and T2 decay compensation: application to carotid artery imaging

  • Author/Authors

    Zhang، نويسنده , , Ling and Kholmovski، نويسنده , , Eugene G. and Guo، نويسنده , , Junyu and Choi، نويسنده , , Seong-Eun Kim and Morrell، نويسنده , , Glen R. and Parker، نويسنده , , Dennis L.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    10
  • From page
    13
  • To page
    22
  • Abstract
    T2-weighted carotid artery images acquired using the turbo spin-echo (TSE) sequence frequently suffer from motion artifacts due to respiration and blood pulsation. The possibility of using HASTE sequence to achieve motion-free carotid images was investigated. The HASTE sequence suffers from severe blurring artifacts due to signal loss in later echoes due to T2 decay. Combining HASTE with parallel acquisition (PHASTE) decreases the number of echoes acquired and thus effectively reduces the blurring artifact caused by T2 relaxation. Further improvement in image sharpness can be achieved by performing T2 decay compensation before reconstructing the PHASTE data. Preliminary results have shown successful suppression of motion artifacts with PHASTE imaging. The image quality was enhanced relative to the original HASTE image, but was still less sharp than a non–motion-corrupted TSE image.
  • Keywords
    HASTE , T2 blurring , Carotid artery imaging , Motion suppression , k-Space inherited parallel acquisition
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2009
  • Journal title
    Magnetic Resonance Imaging
  • Record number