• Title of article

    A volume resolution phantom for MRI

  • Author/Authors

    Moon، نويسنده , , Sang Yun and Hornak، نويسنده , , Joseph P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    4
  • From page
    286
  • To page
    289
  • Abstract
    Multisite quantitative magnetic resonance imaging (qMRI) of volume requires a small isotropic point spread-function (PSF) that is spatially, temporarily, and platform invariant. A phantom which will allow rapid assessment of this metric throughout the imaged volume without repositioning will assist certification of imaging sites for use in qMRI studies based on volume. This paper presents a phantom design for this purpose with a three-dimensional repeating pattern throughout its 800-cm3 volume. The image of the pattern from the phantom contains a series of positive signal points and lines which can be used to measure the PSF, gradient linearity, gradient orthogonality, and B0 homogeneity at multiple locations throughout its volume. The phantom is readily constructed, can be filled with any nuclear magnetic resonance signal-bearing liquid, and the design is scalable to cover larger volumes.
  • Keywords
    linearity , RESOLUTION , Point spread function , Magnetic Resonance Imaging , Resolution phantom , MRI , Qualitative MRI , qMRI
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2010
  • Journal title
    Magnetic Resonance Imaging
  • Record number