• Title of article

    Assessment of MRI issues for the Argus II Retinal Prosthesis

  • Author/Authors

    Weiland، نويسنده , , James D. and Faraji، نويسنده , , Boozarjomehr and Greenberg، نويسنده , , Robert J. and Humayun، نويسنده , , Mark S. and Shellock، نويسنده , , Frank G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    8
  • From page
    382
  • To page
    389
  • Abstract
    Objective jective was to evaluate magnetic resonance imaging (MRI) issues (magnetic field interactions, heating, artifacts and functional alterations) at 1.5 T and 3 T for the Argus II Retinal Prosthesis (Second Sight Medical Products, Sylmar, CA, USA). als and Methods rdized protocols were used to assess magnetic field interactions (translational attraction and torque; 3 T, worst case), MRI-related heating (1.5 and 3 T), artifacts (3 T; worst case) and functional changes (1.5 and 3 T) associated with MRI. s gnetic field interactions were acceptable. MRI-related heating, which was studied at a relatively high, MR system-reported whole body averaged specific absorption rates, will not pose a hazard to the patient under the conditions used for testing. While artifacts were “moderate” in relation to the dimensions of the Argus II Retinal Prosthesis, optimization of MRI parameters can reduce the size of the artifacts. Exposures to MRI conditions at 1.5 and 3 T did not damage or alter the functional aspects of the Argus II Retinal Prosthesis. sions sideration of the test results, a patient with the Argus II Retinal Prosthesis may undergo MRI at 1.5 T or 3 T when specific guidelines and MRI conditions are followed, including those advised by the manufacturer.
  • Keywords
    Magnetic Resonance Imaging , retinal prosthesis , MRI , MRI , artifacts , implants , Safety
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2012
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1833275