Title of article :
Improvement of image quality using BLADE sequences in brain MR imaging
Author/Authors :
E. Lavdas، نويسنده , , Eleftherios and Mavroidis، نويسنده , , Panayiotis and Kostopoulos، نويسنده , , Spiros and Glotsos، نويسنده , , Dimitrios and Roka، نويسنده , , Violeta and Topalzikis، نويسنده , , Theofilos and Bakas، نويسنده , , Athanasios and Oikonomou، نويسنده , , Georgia and Papanikolaou، نويسنده , , Nikos and Batsikas، نويسنده , , Georgios and Kaffes، نويسنده , , Ioannis and Kechagias، نويسنده , , Dimitrios، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
12
From page :
189
To page :
200
Abstract :
The purpose of this study is to compare two types of sequences in brain magnetic resonance (MR) examinations of uncooperative and cooperative patients. For each group of patients, the pairs of sequences that were compared were two T2-weighted (T2-W) fluid attenuated inversion recovery sequences with different k-space trajectories (conventional Cartesian and BLADE) and two T2-TSE weighted with different k-space trajectories (conventional Cartesian and BLADE). Twenty-three consecutive uncooperative patients and 44 cooperative patients, who routinely underwent brain MR imaging examination, participated in the study. Both qualitative and quantitative analyses were performed based on the signal-to-noise ratio, contrast-to-noise ratio (CNR), and relative contrast (ReCon) measures of normal anatomic structures. The qualitative analysis was performed by experienced radiologists. Also, the presence of motion, other (e.g., Gibbs, susceptibility artifacts, phase encoding from vessels) artifacts and pulsatile flow artifacts was evaluated. uncooperative group of patients, BLADE sequences were superior to the corresponding conventional sequences in all the cases. Furthermore, the differences were found to be statistically significant in almost all the cases. In the cooperative group of patients, BLADE sequences were superior to the conventional sequences with the differences of the CNR and ReCon values in nine cases being statistically significant. Furthermore, BLADE sequences eliminated motion and other artifacts and T2 FLAIR BLADE sequences eliminated pulsatile flow artifacts. sequences (T2-TSE and T2 FLAIR) should be used in brain MR examinations of uncooperative patients. In cooperative patients, T2-TSE BLADE sequences may be used as part of the routine protocol and orbital examinations. T2 FLAIR BLADE sequences may be used optionally in examinations of AVM, orbits, haemorrhages, ventricular lesions, lesions in the frontal lobe, periventricular lesions, lesions in regions close to artifacts and lesions in posterior fossa.
Keywords :
1.5T MRI , Motion artifacts , BLADE sequences , Brain examination
Journal title :
Magnetic Resonance Imaging
Serial Year :
2013
Journal title :
Magnetic Resonance Imaging
Record number :
1833421
Link To Document :
بازگشت