    Parallel EPI artifact correction (PEAC) for N/2 ghost suppression in neuroimaging applications

  • Author/Authors

    Li، نويسنده , , Hailong and Fox-Neff، نويسنده , , Kristen and Vaughan، نويسنده , , Benjamin and French، نويسنده , , Donald and Szaflarski، نويسنده , , Jerzy P. and Li، نويسنده , , Yu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    1022
  • To page
    1028
  • Abstract
    Echo Planar Imaging (EPI) is a neuroimaging tool for clinical practice and research investigation. Due to odd-even echo phase inconsistencies, however, EPI suffers from Nyquist N/2 ghost artifacts. In standard neuroimaging protocols, EPI artifacts are suppressed using phase correction techniques that require reference data collected from a reference scan. Because reference-scan based techniques are sensitive to subject motion, EPI performance is sub-optimal in neuroimaging applications. In this technical note, we present a novel EPI data processing technique which we call Parallel EPI Artifact Correction (PEAC). By introducing an implicit data constraint associated with multi-coil sensitivity in parallel imaging, PEAC converts phase correction into a constrained problem that can be resolved using an iterative algorithm. This enables “reference-less” EPI that can improve neuroimaging performance. In the presented work, PEAC is investigated using a standard functional magnetic resonance imaging (fMRI) protocol with multi-slice 2D EPI. It is demonstrated that PEAC can suppress ghost artifacts as effectively as the standard reference-scan based phase correction technique used on a clinical MRI system. We also found that PEAC can achieve dynamic phase correction when motion occurs.
  • Keywords
    MRI , EPI , Nyquist N/2 ghost , Grappa
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2013
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1833541