Title of article :
An isocenter position verification device for electronic portal imaging: physical and dosimetrical characteristics
Author/Authors :
Claus، نويسنده , , Y. De Deene، نويسنده , , Y and Remouchamps، نويسنده , , V and De Wagter، نويسنده , , C and Van Vaerenbergh، نويسنده , , K and Schauvliege، نويسنده , , Els Van Duyse، نويسنده , , B and Speleers، نويسنده , , B and Duthoy، نويسنده , , W and De Neve، نويسنده , , W، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
7
From page :
759
To page :
765
Abstract :
The physical and dosimetrical characteristics of a device, designed to visualize the isocenter position on electronic portal images, were examined. The device, to be mounted on the gantry head of the accelerator, containing five spheric lead markers, was designed in order to visualize the isocenter position on portal images. A quality control device was designed to check the reliability of this technique. The disturbance of the dose distribution by the markers was studied with gel dosimetry. The use of markers resulted in a precise and accurate method to visualize the isocenter on portal images. A maximum underdosage of 11%, due to attenuation by the markers, was observed. The use of markers to visualize the isocenter position on portal images, is a fast and reliable method when analyzing patient setup errors with online electronic portal imaging.
Keywords :
electronic portal imaging , isocenter , imagerie portale électronique , Marker , marqueur , isocentre
Journal title :
Cancer Radiotherapie
Serial Year :
2001
Journal title :
Cancer Radiotherapie
Record number :
1841323
Link To Document :
بازگشت