• Title of article

    Severe aortic insufficiency secondary to an aortic bioprosthesis tear

  • Author/Authors

    Beauchesne، نويسنده , , Luc M and Veinot، نويسنده , , John D. and Higginson، نويسنده , , Lyall A and Mesana، نويسنده , , Thierry، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    3
  • From page
    165
  • To page
    167
  • Abstract
    A patient with a 10-year-old Medtronic Hancock II porcine aortic bioprosthesis developed severe aortic insufficiency. A transesophageal echocardiogram showed a long and mobile mass attached to the bioprosthesis which was consistent with a torn cusp. The patient underwent replacement of the prosthesis with a mechanical valve. Pathological examination showed two subacute tears arising from the same suture buttressing site. These two tears allowed a portion of the valve apparatus to prolapse.
  • Keywords
    Prosthetic valve , Bioprosthesis , Aortic insufficiency
  • Journal title
    Cardiovascular Pathology
  • Serial Year
    2004
  • Journal title
    Cardiovascular Pathology
  • Record number