Title of article
Severe aortic insufficiency secondary to an aortic bioprosthesis tear
Author/Authors
Beauchesne، نويسنده , , Luc M and Veinot، نويسنده , , John D. and Higginson، نويسنده , , Lyall A and Mesana، نويسنده , , Thierry، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
3
From page
165
To page
167
Abstract
A patient with a 10-year-old Medtronic Hancock II porcine aortic bioprosthesis developed severe aortic insufficiency. A transesophageal echocardiogram showed a long and mobile mass attached to the bioprosthesis which was consistent with a torn cusp. The patient underwent replacement of the prosthesis with a mechanical valve. Pathological examination showed two subacute tears arising from the same suture buttressing site. These two tears allowed a portion of the valve apparatus to prolapse.
Keywords
Prosthetic valve , Bioprosthesis , Aortic insufficiency