• Title of article

    Toward a final design for the Birmingham boron neutron capture therapy neutron beam

  • Author/Authors

    Beynon، T. D. نويسنده , , Green، S. نويسنده , , Allen، D. A. نويسنده , , James، N. D. نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2017
  • Pages
    -76
  • From page
    77
  • To page
    0
  • Abstract
    A preliminary study of the physical characteristics of a miniature multileaf collimator (mMLC) used with 4 MV x rays is reported. The mMLC attached to the accessory mount of a class C or D Varian linear accelerator (Varian Oncology Systems, Palo Alto, CA) with a source to aperture distance of 65 cm. The held penumbra using the small leaves was found to be consistent with the anticipated field penumbra using photon jaws at the same source to aperture distance as the mMLC. The percentage depth dose values of square fields were found to be consistent with the fields collimated with the upper and lower jaws. Output factors for the very small fields were found to vary rapidly. Circular fields could be produced with depth dose characteristics similar to those produced using conical tertiary collimators, commonly used for radiosurgery, but with a broader penumbra
  • Keywords
    BNCT , accelerator , Monte Carlo , epithermal neutrons , MCNP
  • Journal title
    MEDICAL PHYSICS
  • Serial Year
    1999
  • Journal title
    MEDICAL PHYSICS
  • Record number

    1845