Évaluation d’une nouvelle table, Exact® IGRT (Varian), sur l’atténuation du faisceau de traitement et la qualité des images
Author/Authors :
Marguet، نويسنده , , M. and Garcia، نويسنده , , R. and Bordage، نويسنده , , M.-C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
8
From page :
42
To page :
49
Abstract :
Purpose couch top designed for IGRT procedures, Exact® IGRT (EIC), has been developped. It has the advantage of having no removable support bars and is thicker than a standard table (Exact®(EC)) in order to create the appropriate rigidity. The aim of this study was to quantify the beam attenuation caused by the EIC and to evaluate the EIC contribution to on-board image quality improvement. als and methods eatment beam attenuation and the surface-dose of the patient were measured for simple radiation geometries at different gantry angles and for two clinical cases, head and neck and posterior pelvis. The 2D and 3D image quality was analysed. s am attenuation at 215° was 2.0% to 4.1% for 6 MV and 1.2% to 2.2% for 18 MV according to the EIC part. The dose-surface tripled between the measurement without table and with table into the beam (similar results between EIC and EC). The EIC influence was significative for the posterior pelvic tumor, acceptable for the head and neck case. The EIC showed a clear improvement of the low contrast resolution (visualisation of a 7 mm and 1% contrast disk [15 mm for the EC]). sion er to take into account the beam attenuation, the EIC should be modeled in the treatment planning system. However, the good quality of 3D images obtained with the EIC allows us to easily visualize soft tissues.
Keywords :
Table de traitement , Qualité d’image , Radiothérapie guidée par l’image (IGRT) , Atténuation par la table , Treatment table , image quality , Image-guided radiotherapy (IGRT) , Table attenuation
Journal title :
Cancer Radiotherapie
Serial Year :
2010
Journal title :
Cancer Radiotherapie
Record number :
1846762
Link To Document :
بازگشت