Title of article :
Determination of the optimal statistical uncertainty to perform electron-beam Monte Carlo absorbed dose estimation in the target volume
Author/Authors :
Isambert، نويسنده , , A. and Brualla، نويسنده , , L. and Benkebil، نويسنده , , M. and Lefkopoulos، نويسنده , , D.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
Purpose of study
Carlo based treatment planning system are known to be more accurate than analytical methods for performing absorbed dose estimation, particularly in and near heterogeneities. However, the required computation time can still be an issue. The present study focused on the determination of the optimum statistical uncertainty in order to minimise computation time while keeping the reliability of the absorbed dose estimation in treatments planned with electron-beams.
als and methods
radiotherapy plans (medulloblastoma, breast and gynaecological) were used to investigate the influence of the statistical uncertainty of the absorbed dose on the target volume dose–volume histograms (spinal cord, intramammary nodes and pelvic lymph nodes, respectively).
s
udy of the dose–volume histograms showed that for statistical uncertainty levels (1 S.D.) above 2 to 3%, the standard deviation of the mean dose in the target volume calculated from the dose–volume histograms increases by at least 6%, reflecting the gradual flattening of the dose–volume histograms.
sions
ork suggests that, in clinical context, Monte Carlo based absorbed dose estimations should be performed with a maximum statistical uncertainty of 2 to 3%.
Keywords :
Monte Carlo , Incertitude statistique , Faisceaux d’électrons , Electron-beams , statistical uncertainty , Monte Carlo
Journal title :
Cancer Radiotherapie
Journal title :
Cancer Radiotherapie