Title of article :
In vitro and finite element analysis of glenoid bone/baseplate interaction in the reverse shoulder design
Author/Authors :
Virani، نويسنده , , Nazeem A. and Harman، نويسنده , , Melinda and Li، نويسنده , , Ke and Levy، نويسنده , , Jonathan and Pupello، نويسنده , , Derek R. and Frankle، نويسنده , , Mark A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
We developed biomechanical and finite element models, using high-strength polyurethane foam blocks, to represent the glenoid bone/baseplate junction to determine if increasing the distance between the glenoid bone and the center of rotation of the glenosphere increases baseplate motion during static loading in the reverse shoulder design. Although there was a general trend toward increased baseplate motion with increasing distance from the glenoid to the center of rotation, in vitro mechanical testing revealed no significant difference between the 7 glenosphere types tested, with average baseplate motion during 1000 load cycles ranging from 90 μm to 120 μm. Results from the finite element analysis strongly correlated with the in vitro mechanical testing. The magnitude of baseplate motion occurring in a modeled representation of bone under simulated physiologic loading conditions was similar for the 7 reverse shoulder glenoid components tested in this study.
Journal title :
Journal of Shoulder and Elbow Surgery
Journal title :
Journal of Shoulder and Elbow Surgery