• Title of article

    Loss of rotator cuff tendon-to-bone interface pressure after reattachment using a suture anchor

  • Author/Authors

    Brassart، نويسنده , , Nicolas and Sanghavi، نويسنده , , Sanjay and Hansen، نويسنده , , Ulrich N. and Emery، نويسنده , , Roger J. and Amis، نويسنده , , Andrew A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    784
  • To page
    789
  • Abstract
    The purpose of this study was to examine the tendon-to-bone interface pressure, contact area, and force after reattaching a tendon to bone by use of a suture and suture anchor. Repairs were made in 8 ovine shoulders in vitro, by use of 3 suture types in each: Ethibond, polydioxanone, or Orthocord. A Tekscan pressure sensor was placed between the tendon and bone and monitored for 1 hour after the repair. The principal finding was a significant loss of approximately 60% of the contact parameters immediately after the suture was tied, followed by further significant loss over the next hour to a mean of only 14% of the initial readings. We concluded that pressure measurement systems that only record the initial maximum pressure would yield overly optimistic results for the actual repair pressure after the repair is completed. The Tekscan system, however, allowed us to monitor pressure reductions that occurred both during and after the repair.
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Serial Year
    2008
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Record number

    1867916