Title of article :
Glenoid implant orientation and cement failure in total shoulder arthroplasty: a finite element analysis
Author/Authors :
Yongpravat، نويسنده , , Charlie S. Kim، نويسنده , , H. Mike and Gardner، نويسنده , , Thomas R. and Bigliani، نويسنده , , Louis U. and Levine، نويسنده , , William N. and Ahmad، نويسنده , , Christopher S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
Background
imize glenoid implant loosening in total shoulder arthroplasty (TSA), the ideal surgical procedure achieves correction to neutral version, complete implant–bone contact, and bone stock preservation. These goals, however, are not always achievable, and guidelines to prioritize their impact are not well established. The purpose of this study was to investigate how the degree of glenoid correction affects potential cement failure.
s
patient-specific computer models were created for 4 TSA scenarios with different permutations of retroversion correction and implant–bone contact. Two bone models were used: a homogeneous cortical bone model and a heterogeneous cortical-trabecular bone model. A 750-N load was simulated, and cement stress was calculated. The risk of cement mantle fracture was reported as the percentage of cement stress exceeding the material endurance limit.
s
ing the glenoid implant in retroversion resulted in the highest risk of cement fracture in a homogeneous bone model (P < .05). In the heterogeneous bone model, complete correction resulted in the highest risk of failure (P = .0028). A positive correlation (ρ = 0.901) was found between the risk of cement failure and amount of exposed trabecular bone.
sions
orating trabecular bone into the model changed the effect of implant orientation on cement failure. As exposed trabecular bone increased, the risk of cement fracture increased. This may be due to shifting the load-bearing support underneath the cement from cortical bone to trabecular bone.
Keywords :
eccentric reaming , retroversion , Computer simulation , decortication , glenoid wear , cement failure , Finite element analysis , Total shoulder arthroplasty
Journal title :
Journal of Shoulder and Elbow Surgery
Journal title :
Journal of Shoulder and Elbow Surgery