Title of article :
Theoretical analysis on ratiometric fluorescent indicators caused biased estimates of intracellular free calcium concentrations
Author/Authors :
Hu، نويسنده , , Zhiwen and Yu، نويسنده , , Liangdeng and Yu، نويسنده , , Zengliang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
9
From page :
179
To page :
187
Abstract :
Ratiometric fluorescent calcium indicator dyes have been widely used for the study of the role of Ca2+ in cell physiopathology. Although these ratiometric dyes offer several advantages over others, they suffer some drawbacks which cause serious errors in measurement of intracellular Ca2+ concentration, [Ca2+]i. The present study systematically analyzes theoretical reasons and technical sources of discrepancies occurring in the measurement of the characteristics of the agonists-induced cells [Ca2+]i. In order to avoid the errors and achieve the accurate determination of [Ca2+]i, this study proposes solutions and suggests some critical measures in both theoretical and technical aspects. Therefore, this analysis can be a valuable tool in clarifying proper usages of fluorescent dyes for [Ca2+]i measurements.
Keywords :
Ratiometric fluorescent indicator , Intracellular free calcium , Dye extrusion , Calibration
Journal title :
Journal of Photochemistry and Photobiology B:Biology
Serial Year :
2005
Journal title :
Journal of Photochemistry and Photobiology B:Biology
Record number :
1874907
Link To Document :
بازگشت