• Title of article

    Optimal size of self-adhesive preapplied electrode pads in pediatric defibrillation

  • Author/Authors

    Samson، نويسنده , , Ricardo A. and Atkins، نويسنده , , Dianne L. and Kerber، نويسنده , , Richard E.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    2
  • From page
    544
  • To page
    545
  • Abstract
    We conclude that to minimize TTI, preapplied self-adhesive electrode pads should be at least 5.8 cm in diameter when used for pediatric patients. The 5.8 cm pad was easily applied to infants. Older children could easily accommodate the 7 cm pad. Therefore, we recommend a pad diameter of 5.8 cm for infants < 1 year, and a pad diameter of 7 cm for older children.
  • Journal title
    American Journal of Cardiology
  • Serial Year
    1995
  • Journal title
    American Journal of Cardiology
  • Record number

    1880673