Title of article
Optimal size of self-adhesive preapplied electrode pads in pediatric defibrillation
Author/Authors
Samson، نويسنده , , Ricardo A. and Atkins، نويسنده , , Dianne L. and Kerber، نويسنده , , Richard E.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
2
From page
544
To page
545
Abstract
We conclude that to minimize TTI, preapplied self-adhesive electrode pads should be at least 5.8 cm in diameter when used for pediatric patients. The 5.8 cm pad was easily applied to infants. Older children could easily accommodate the 7 cm pad. Therefore, we recommend a pad diameter of 5.8 cm for infants < 1 year, and a pad diameter of 7 cm for older children.