Author/Authors :
Zaim، نويسنده , , Bulent R. and Zaim، نويسنده , , Sina H. and Rankin، نويسنده , , Andrew C. and McGovern، نويسنده , , Brian A. and Garan، نويسنده , , Hasan and Ruskin، نويسنده , , Jeremy N.، نويسنده ,
Abstract :
We propose that in clinical practice, whenever possible, the VT detection interval should be selected by adding ≥60 ms to the induced maximal VT cycle length in order to ensure a high sensitivity for the detection of future spontaneous VT episodes.