Low-Energy Endocardial Defibrillation Using Dual, Triple, and Quadruple Electrode Systems
Author/Authors
Munsif، نويسنده , , Anand N and Saksena، نويسنده , , Sanjeev and DeGroot، نويسنده , , Paul and Krol، نويسنده , , Ryszard B. and Mathew، نويسنده , , Philip and Giorgberidze، نويسنده , , Irakli and Kaushik، نويسنده , , Raj R and Mehra، نويسنده , , Rahul، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
8
From page
1632
To page
1639
Abstract
The feasibility of achieving both universal application of nonthoracotomy leads and low (≤15 J) defibrillation energy requirements by optimizing lead system configuration for use with low-output (<30 J) biphasic shock pulse generators was examined. Sixteen patients (mean age 62 ± 8 years and mean left ventricular ejection fraction of 38 ± 15%) were included in the study. All patients had either experienced syncope with induced ventricular tachycardia (n = 4) or had documented sustained ventricular tachycardia (n = 7) or ventricular fibrillation (n = 5). Defibrillation threshold testing was performed in 2 stages on different days in these patients. In the first stage, 2 defibrillation catheter electrodes were positioned in the right ventricle and superior vena cava with an axillary cutaneous patch. Fifteen-joule, 10- and 5-J biphasic shocks were delivered across 3 different electrode configurations—right ventricle to superior vena cava, right ventricle to axillary patch, right ventricle to a combination of superior vena cava and axillary patch. In the second stage, an 80-ml can electrode was added subcutaneously in a pectoral location to the previous leads. Configurations compared were the right ventricle to pectoral can, and right ventricle to an “array”-combining superior vena cava, can, and axillary patch leads. The defibrillation threshold was determined using a step-down method. In stage 1, mean defibrillation threshold for the right ventricle to axillary patch (12.7 ± 5.9 J) and right ventricle to superior vena cava plus axillary patch (9.8 ± 5.2 J) configurations was lower than the right ventricle to superior vena cava configuration (14.2 ± 6.4 J, p <0.05). In stage 2, the defibrillation was higher for the right ventricle to pectoral can (9.2 ± 5.1 J) configuration compared with the right ventricle to the array (5.6 ± 3.6 J, p ≤0.05). The right ventricle to array had the lowest defibrillation threshold, whereas the right ventricle to pectoral can was the best dual electrode system. Low-energy endocardial defibrillation (≤10 J) was feasible in 72% of tested patients with >1 electrode configuration at 10 J, whereas only 53% of successful patients could be reverted at >1 electrode configuration at 5 J (p <0.05). Reduction in maximum pulse generator output to ≤25 J using these electrode configurations with bidirectional shocks is feasible and maintains an adequate safety margin.
tudy examines the feasibility of achieving both universal application of nonthoracotomy leads and low (≤15 J) defibrillation energy requirements by optimizing 5 different lead system configurations. Reduction in mean defibrillation threshold to 5.6 ± 3.6 J could be observed, which allowed pulse generators with a maximum output of ≤25 J and yet maintained an adequate safety margin in all patients.
Journal title
American Journal of Cardiology
Serial Year
1997
Journal title
American Journal of Cardiology
Record number
1885079
Link To Document