Title of article :
Clinical significance of multiple sensor options: rate response optimization, sensor blending, and trending
Author/Authors :
Clémenty، نويسنده , , Jacques and Barold، نويسنده , , Serge S and Garrigue، نويسنده , , Stéphane and Shah، نويسنده , , Dipen C and Jaïs، نويسنده , , Pierre and Le Métayer، نويسنده , , Philippe and Haissaguerre، نويسنده , , Michel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
166
To page :
171
Abstract :
The gold standard for rate modulation is the sinus node. To improve the rate modulation provided by artificial sensors, new sensors have to be developed or 2 different sensor systems can be combined within a single device. Association combination of a sensor with a rapid-response fast-rate increase sensor (activity) and a progressive, more specific sensor (QT ventilation) is generally used. Sensor combinations require adequate sensor blending for signal production and prioritization during rate modulation. However, in the new devices, some other aspects of rate modulation could be taken into consideration, particularly circadian rate variations to obtain lower rates at nighttime than during daytime, and automatic adaptation of the slope of rate increase during exercise, according to the patient’s fitness, heart function, age, etc. Despite the need for automaticity, manual programming could continue to be useful to adapt rate modulation with data from sensor trending memories.
Journal title :
American Journal of Cardiology
Serial Year :
1999
Journal title :
American Journal of Cardiology
Record number :
1889486
Link To Document :