Title of article :
Efficacy of atrial antitachycardia pacing using the Medtronic AT500 pacemaker in patients with congenital heart disease
Author/Authors :
Stephenson، نويسنده , , Elizabeth A. and Casavant، نويسنده , , David and Tuzi، نويسنده , , Joann and Alexander، نويسنده , , Mark E. and Law، نويسنده , , Ian and Serwer، نويسنده , , Gerald and Strieper، نويسنده , , Margaret and Walsh، نويسنده , , Edward P. and Berul، نويسنده , , Charles I.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
871
To page :
876
Abstract :
Patients with congenital heart disease are vulnerable to atrial tachyarrhythmias, especially after atrial surgeries. We evaluated the efficacy of atrial arrhythmia detection and antitachycardia pacing (ATP) using the Medtronic AT500 pacemaker in 28 patients with congenital heart disease (age 30 ± 18 years). Of 15 patients with atrial arrhythmias, 14 had atrial tachycardia events that were appropriately detected. ATP was enabled for 167 treatable episodes, successfully converting 90 (54%). Rhythms classified as ventricular tachycardia were detected 127 times, yet most were actually atrial or sinus tachycardia with 1:1 atrioventricular conduction. Atrial tachycardias in congenital heart disease are amenable to ATP alogorithms in the AT500 pacemaker.
Journal title :
American Journal of Cardiology
Serial Year :
2003
Journal title :
American Journal of Cardiology
Record number :
1896486
Link To Document :
بازگشت