Title of article :
Incidence of thrombus formation on the CardioSEAL and the Amplatzer interatrial closure devices
Author/Authors :
Anzai، نويسنده , , Hitoshi and Child، نويسنده , , John and Natterson، نويسنده , , Barbara and Krivokapich، نويسنده , , Janine and Fishbein، نويسنده , , Michael C and Chan، نويسنده , , Vicki K and Tobis، نويسنده , , Jonathan M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
6
From page :
426
To page :
431
Abstract :
Transcatheter closure for atrial septal defect (ASD) and patent foramen ovale (PFO) is a promising alternative to surgical closure or anticoagulant therapy. A potential complication is thrombus formation on the device after implantation. From February 2001 to June 2003, 66 patients with atrial communication were treated successfully with the Amplatzer device (16 septal and 20 PFO occluders) or the CardioSEAL device (30). Patients were discharged on antiplatelet medication (aspirin and clopidogrel) and/or anticoagulation. Fifty patients (76%) had transesophageal echocardiography (TEE) 1 month after device implantation (28 ± 10 days). No patient experienced a thromboembolic episode during follow-up. TEE revealed that thrombus formation occurred more frequently on the CardioSEAL device (5 of 23 patients; 22%) than on the Amplatzer device (0 of 27 patients; 0%) (p = 0.02). Although thrombus disappeared or markedly diminished after additional anticoagulation therapy in 3 patients, 1 patient had surgical explantation of the device due to progressive increase in the size of thrombus with hypermobility despite intensive anticoagulation therapy. There was no variable associated with the presence of thrombus formation on the occluder other than the use of the CardioSEAL device. One month after insertion, the CardioSEAL device is more likely to have thrombus present than the Amplatzer device.
Journal title :
American Journal of Cardiology
Serial Year :
2004
Journal title :
American Journal of Cardiology
Record number :
1897051
Link To Document :
بازگشت