Abstract :
The accuracy of a hand-carried cardiac ultrasound (HCCU) device for the diagnosis of valvular regurgitations using color-flow Doppler was assessed. One hundred twenty hospitalized patients with at least mild valvular regurgitation by standard echocardiography were evaluated by 2 cardiologists using HCCU. The HCCU operator identified 99% of clinically significant valvular regurgitations and assessed the severity correctly in 83% of cases. For mild regurgitation, the HCCU sensitivity and specificity were 82% and 93%, respectively, resulting in correct assessments of severity in 71% of cases. HCCU used by trained cardiologists has high sensitivity for the detection and accurate assessment of the severity of clinically relevant valvular regurgitation.