Kirvaitis، نويسنده , , Romas J. and Heuser، نويسنده , , Richard R. and Das، نويسنده , , Tony S. and Laster، نويسنده , , Steven B. and Dippel، نويسنده , , Eric J. and Gammon، نويسنده , , Roger S. and Botti Jr.، نويسنده , , Charles F. and Murphy، نويسنده , , Bruce E. and Biggs، نويسنده , , Thomas A. and Shimshak، نويسنده , , Thomas A. and Laird، نويسنده , , John R. and Foster، نويسنده , , Malcolm T، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
4
From page :
1081
To page :
1084
Abstract :
Optical coherent reflectometry, a forward-looking, fiberoptic-guided device was used in 72 patients to direct radiofrequency energy across the central intraluminal portion of 75 chronic total occlusions in peripheral arteries (iliac, femoral, and popliteal) that failed attempts with conventional guidewires. The system was successful in crossing 76% of the chronic total occlusions with no clinical perforations or distal embolizations, and complications consisted of a single dissection greater than or equal to grade C.