Ultrasonic Biophysical Measurements in the Normal Human Fetus for Optimal Design of the Monolithic Fetal Pacemaker
Author/Authors :
Jocelyne Fayn، نويسنده , , Evgueni and Chou، نويسنده , , Howard A. and Park، نويسنده , , DaeGyun and Zavitz، نويسنده , , Daniel H. and Cuneo، نويسنده , , Bettina F. and Mahan، نويسنده , , Vicki L. and Guleçyuz، نويسنده , , Mehmet and Curran، نويسنده , , Letitia and Lipson، نويسنده , , David and Quillen Jr.، نويسنده , , Edmond W. and Petrikovsky، نويسنده , , Boris M. and Ovadia، نويسنده , , Marc، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
4
From page :
1267
To page :
1270
Abstract :
Ultrasound measurements, including xiphoid-to-pericardial distance and deployment angle, were made on human fetuses as a function of gestational age for the purpose of assessing the likelihood of 3 failure modes of a monolithic fetal pacemaker, including primary positioning failure due to device length and secondary dislodgement failure due to somatic growth. The small variation of the measurements over the gestational age range relevant to device implantation for the major indications of the device (for complete heart block complicated by hydrops and for bradycardia risk after fetal surgery or intrauterine intervention) predicts a small likelihood of these failure modes.