Title of article :
Phase-shift cavity ring-down spectroscopy to determine absolute line intensities
Author/Authors :
van Helden، نويسنده , , J.H. and Schram، نويسنده , , D.C. and Engeln، نويسنده , , R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
Cavity ring-down detection techniques can sensitively determine frequency-dependent absorption cross-sections of gasses. However, so-called line-width problems and amplified spontaneous emission of the laser light source lowers the technique’s quantitative accuracy. Using phase-shift cavity ring-down spectroscopy (PSCRD), we measured absolute line intensities of the spin-forbidden transitions in the b 1 Σ g + ( ν ′ = 0 ) ← X 3 Σ g - ( ν ″ = 0 ) band of molecular oxygen. Our results were within 4% of values obtained from the HITRAN database, demonstrating the accuracy of PSCRD, when corrected for amplified spontaneous emission. Its high sensitivity (2 × 10−8 cm−1), simplicity and high duty cycle make PSCRD a powerful diagnostic technique.
Journal title :
Chemical Physics Letters
Journal title :
Chemical Physics Letters