Title of article :
Biosensor-based determination of folic acid in fortified food
Author/Authors :
Bostrِm Caselunghe، نويسنده , , Malin and Lindeberg، نويسنده , , Johan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A newly developed method for the quantification of folic acid in fortified food is presented. An immunoaffinity-based optical biosensor was used to determine folic acid concentration levels in milk powder, infant formula and cereal samples. Accuracy of the method (88–101%) was demonstrated with the analysis of five reference samples. A collaborative precision study, where ten participants at four different laboratories analysed a set of ten samples, resulted in repeatability relative standard deviations of 2–8% and reproducibility relative standard deviations of 4–10%.
Keywords :
Biosensors , Folic Acid , Analyses , Fortified food
Journal title :
Food Chemistry
Journal title :
Food Chemistry