Title of article :
Determination of TOTOX value in palm oleins using a FI-potentiometric analyzer
Author/Authors :
Wai، نويسنده , , Wan Tatt and Saad، نويسنده , , Bahruddin and Lim، نويسنده , , Boey Peng، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
A flow injection analyzer for the determination of total oxidation value (TOTOX) in palm oleins was developed based on the sequential determination of peroxide value (PV) and anisidine value (AV) using a common triiodide ( I 3 − ) potentiometric detector. The PV and AV parameters were sequentially determined by manipulating a switching valve. The mixing of the oil and reagents was effective by applying a short vortex (30 s). The TOTOX value is obtained from the potentiometric readings of PV and AV (TOTOX = 2PV + AV). Using palm olein as standard, the method gives TOTOX value that is linear over 1.7–79.0 with a detection limit of 1.64. The sampling frequency is about 60 h−1. Twenty aged and thermally stressed palm olein samples were analyzed and good correlations (r2 = 0.9852) were found between the proposed FIA and the official methods.
Keywords :
TOTOX value , Oxidative stability , palm olein , Flow injection analysis
Journal title :
Food Chemistry
Journal title :
Food Chemistry