• Title of article

    Determination of melamine by flow injection analysis based on chemiluminescence system

  • Author/Authors

    Zeng، نويسنده , , Hua-jin and Yang، نويسنده , , Ran and Wang، نويسنده , , Qingwen and Li، نويسنده , , Jianjun and Qu، نويسنده , , Ling-bo، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    842
  • To page
    846
  • Abstract
    In this paper, based upon the phenomenon that melamine can obviously enhance the CL signal of the luminol–H2O2 system in basic medium, a simple, rapid and sensitive flow injection chemiluminescence (FI-CL) method for the determination of melamine has been developed. Under the optimum conditions, the linear range for the determination of melamine was 0.2–80 μg mL−1 with a detection limit of 0.12 μg mL−1 calculated as proposed by IUPAC and a relative standard deviation of 3.26% for 11 solutions of 10 μg mL−1 melamine on the same day. The proposed method was satisfactorily applied to determine melamine in milk-based products and satisfactory results were obtained without interferences from the sample matrix. Moreover, one assay produce takes only 25 s and the minimum sampling rate is about 120 samples h−1, which indicated that the FI-CL method was suitable for high throughput and real-time melamine analysis.
  • Keywords
    Melamine , Milk-based products , Flow injection analysis , Chemiluminescence
  • Journal title
    Food Chemistry
  • Serial Year
    2011
  • Journal title
    Food Chemistry
  • Record number

    1965134