Xu، نويسنده , , Zhen-Lin and Deng، نويسنده , , Chih-Hao and Deng، نويسنده , , Xing-Fei and Yang، نويسنده , , Jin-Yi and Jiang، نويسنده , , Yue-Ming and Zeng، نويسنده , , Dao-Ping and Huang، نويسنده , , Fang and Shen، نويسنده , , Yu-Dong and Lei، نويسنده , , Hongtao and Wang، نويسنده , , Hong and Sun، نويسنده , , Yuan-Ming، نويسنده ,
Abstract :
A multi-analytes method for monitoring of organophosphorus pesticides (OPs) using a combination of broad-specificity direct competitive enzyme-linked immunosorbent assay (dcELISA) and high-performance liquid chromatography–tandem mass spectrometry (HPLC–MS/MS) was developed. The reaction formats of dcELISA as well as the matrix effects of vegetable samples by different treatments were studied. The dcELSIA based on horseradish peroxidase-labelled monoclonal antibody and solid-phase extraction can analyse 42 samples in duplicate simultaneously for 12 OPs with a limit of detection at 20 μg L−1 within 40 min, with good accuracy and reproducibility. For screening purpose, the dcELISA can distinguish positive samples from hundreds of negative samples at a rapid, high-throughput and low cost manner. The positive samples can be following confirmed by HPLC–MS/MS for the kinds and the relative amounts of OPs. The method is suitable for monitoring of OP contamination in vegetables samples with high-efficiency and low cost.
Keywords :
Direct competitive ELISA , Organophosphorus pesticides , Matrix effect , Solid-phase extraction , HPLC–MS/MS