Title of article :
Ultrasensitive IL-6 electrochemical immunosensor based on Au nanoparticles-graphene-silica biointerface
Author/Authors :
Wang، نويسنده , , Guangfeng and He، نويسنده , , Xiuping and Chen، نويسنده , , Ling and Zhu، نويسنده , , Yanhong and Zhang، نويسنده , , Xiaojun، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
6
From page :
714
To page :
719
Abstract :
An Interleukin-6 (IL-6) electrochemical immunosensor was fabricated based on the Au nanoparticles (AuNP)-graphene-silica sol–gel as immobilization biointerface and AuNP-polydopamine (PDA)@carbon nanotubes (CNT) as the label of HRP-bound antibodies. The AuNP-graphene-silica sol–gel film was prepared in situ and modified on the ITO electrode, providing a stable network for the immobilization of antibody and exhibiting a dynamic working range of 1–40 pg/mL with a low detection limit of 0.3 pg/mL IL-6 (at 3 s). The results of serum samples with the sensor received an acceptable agreement with the ELISA method. Importantly, this method provided a promising ultrasensitive assay strategy for clinical applications.
Keywords :
Polydopamine , Interleukin-6 , Electrochemical immunosensor , graphene
Journal title :
Colloids and Surfaces B Biointerfaces
Serial Year :
2014
Journal title :
Colloids and Surfaces B Biointerfaces
Record number :
1978319
Link To Document :
بازگشت