Title of article :
Application of Polypyrrole Coated Stainless-Steel Wire to the Headspace Solid-Phase Microextraction of Aliphatic Amines
Author/Authors :
Kaykhaii، M. نويسنده 1Department of Chemistry, Faculty of Sciences, Sistan & Balouchestan University, Zahedan 98135-674, Islamic Republic of Iran , , Saffari، F. نويسنده 2Food, Drug and Hygienic Laboratory, Zahedan University of Medical Sciences, Zahedan, Islamic Republic of Iran ,
Issue Information :
فصلنامه با شماره پیاپی 0 سال 2008
Pages :
7
From page :
111
To page :
117
Abstract :
The electrochemical coating technique was used for the preparation of a polypyrrole coating on a stainless-steel wire, and applied as a fibre for solid phase microextraction (SPME). The polypyrrole fibre was employed for analyzing four volatile aliphatic amines (ethylamine, propylamine, butylamine and pentylamine) in water by headspace SPME using gas chromatography-flame ionization detection (GC-FID). Results showed that this fibre coating is suitable for the successful extraction of the target analytes. The effects of the extraction parameters including exposure time, sampling temperature, salt concentration, and stirring rate on the extraction efficiency was optimized. Relative standard deviations (RSDs) was determined ? 4.3%. The correlation coefficient was satisfactory (r2 > 0.98) for the studied analytes. Detection limits were between 0.9 and 1.6 µg.L?1. Good recoveries (99-106 %) were obtained when spiked water and wastewater samples were analyzed.
Journal title :
Journal of Sciences
Serial Year :
2008
Journal title :
Journal of Sciences
Record number :
1983794
Link To Document :
بازگشت