Author/Authors :
Kazkaz، نويسنده , , K. and Foxe، نويسنده , , M. and Bernstein، نويسنده , , A. and Hagmann، نويسنده , , C. and Jovanovic، نويسنده , , I. and Sorensen، نويسنده , , P. and Stoeffl، نويسنده , , W.S. and Winant، نويسنده , , C.D.، نويسنده ,
Abstract :
We have built a gas-phase argon ionization detector to measure small nuclear recoil energies ( < 10 keVee ) . In this paper, we describe the detector response to X-ray and gamma calibration sources, including analysis of pulse shapes, software triggers, optimization of gas content, and energy- and position-dependence of the signal. We compare our experimental results against simulation using a 5.9-keV X-ray source, as well as higher-energy gamma sources up to 1332 keV. We conclude with a description of the detector, DAQ, and software settings optimized for a measurement of the low-energy nuclear quenching factor in gaseous argon. This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory in part under Contract W-7405-Eng-48 and in part under Contract DE-AC52-07NA27344. Funded by Lab-wide LDRD. LLNL-JRNL-415990-DRAFT.
Keywords :
Nuclear quench factor , Gas scintillation proportional counter , argon , Anti-neutrino detector