Title of article
System overview of liquid xenon calorimeter for the MEG experiment
Author/Authors
Iwamoto، نويسنده , , Toshiyuki، نويسنده ,
Pages
4
From page
92
To page
95
Abstract
MEG experiment aims to search for lepton flavor violating μ + → e + γ decay with a sensitivity of 10 - 13 branching ratio. To detect 52.8 MeV γ ʹs , we have been developing a new photon detector with 900 L of liquid xenon viewed by 8462 in. photomultiplier tubes. Construction of the detector was finished in 2007, and the methods to keep liquid xenon stably were confirmed during the engineering run. In 2008, a calibration was done by using 55 MeV γ -rays from π 0 decays to estimate the detector performance. In autumn, MEG experiment has started taking physics data. The detector stability had been monitored by using 17.6 MeV γ ʹs from Li ( p , γ ) Be reaction, etc. I will summarize here the detector preparation, operation, calibration, and monitoring methods, mainly concentrating on the hardware related things.
Keywords
Liquid xenon , Calorimeter , MEG experiment , Lepton flavor violation
Journal title
Astroparticle Physics
Record number
1992071
Link To Document