Author/Authors :
Gervino، نويسنده , , G. W. Marino، نويسنده , , C. and Silvestri، نويسنده , , F. and Lavagno، نويسنده , , A. and Truc، نويسنده , , F.، نويسنده ,
Abstract :
In this paper we present the dosimetry analysis in terms of stability and repeatability of the signal and dose rate dependence of a synthetic single crystal diamond grown by Chemical Vapor Deposition (CVD) technique. The measurements carried out by 5 MeV X-ray photons beam show very promising results, even if the dose rate detector response points out that the charge trapping centers distribution is not uniform inside the crystal volume. This handicap that affects the detectors performances, must be ascribed to the growing process. Synthetic single crystal diamonds could be a valuable alternative to air ionization chambers for quality beam control and for intensity modulated radiation therapy beams dosimetry.
Keywords :
dosimetry , X-rays , ? - Rays , diamond