• Title of article

    Well Ge and semi-planar Ge (HP) detectors for low-level gamma-spectrometry

  • Author/Authors

    F. El-Daoushy، نويسنده , , F. and Garcia-Tenorio، نويسنده , , R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    9
  • From page
    376
  • To page
    384
  • Abstract
    Two gamma spectrometers, a well Ge and a semi-planar Ge (HP), are now installed and calibrated for measurements of environmental radioactivity with special attention being paid to the requirements needed for aquatic mass-balance studies, including isotope-particle dynamics and dating of recent deposits. These facilities will complement the previously developed isotope dilution alpha spectrometry (PIPS detectors) and radon emanation (ionization chambers) techniques for 210Pb dating. A careful evaluation of efficiencies and background of the Ge detectors, together with the influence of the sampleʹs intrinsic activity on the lower limit of detection (LLD), allowed us to study the relative importance of detector and shield specifications on low level gamma spectrometry of environmental samples with various sizes and composition.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1995
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    1993583