Title of article
Comparison of n-γ discrimination by zero-crossing and digital charge comparison methods
Author/Authors
Wolski، نويسنده , , D and Moszy?ski، نويسنده , , M and Ludziejewski، نويسنده , , T and Johnson، نويسنده , , A and Klamra، نويسنده , , W and Skeppstedt، نويسنده , , ?، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
9
From page
584
To page
592
Abstract
A comparative study of the n-γ discrimination done by the digital charge comparison and zero-crossing methods was carried out for a 130 mm in diameter and 130 mm high BC501A liquid scintillator coupled to a 130 mm diameter XP4512B photomultiplier. The high quality of the tested detector was reflected in a photoelectron yield of 2300 ± 100 phe/MeV and excellent n-γ discrimination properties with energy discrimination thresholds corresponding to very low neutron (or electron) energies. The superiority of the ZC method was demonstrated for the n-γ discrimination method alone, as well as, for the simultaneous separation by the pulse shape discrimination and the time-of-flight methods down to about 30 keV recoil electron energy. The digital charge comparison method fails for a large dynamic range of energy and its separation is weakly improved by time-of-flight method for low energies.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1995
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
1994957
Link To Document