Author/Authors :
Schنfer، نويسنده , , W. and Jansen، نويسنده , , E. and Skowronek، نويسنده , , R. and Will، نويسنده , , G. and Kockelmann، نويسنده , , W. and Schmidt، نويسنده , , W. and Tietze-Jaensch، نويسنده , , H.، نويسنده ,
Abstract :
The ROTAX instrument at ISIS is run as an angle-dispersive time-of-flight instrument using the JULIOS scintillation detector in forward-, 90°- and backscattering geometry; d-spacing coverage is from more than 50 Å to less than 0.2 Å. Δd/d-resolution is calculated to 3.5 × 10−3 in backscattering. Tests and examples of a multipurpose use of the neutron instrument as powder diffractometer for crystallographic and magnetic structure analysis are presented; the additional use as texture diffractometer is discussed.