Author/Authors :
Zanevsky، نويسنده , , Yu.V and Chernenko، نويسنده , , S.P and Cheremukhina، نويسنده , , G.A and Fateev، نويسنده , , O.V and Kheiker، نويسنده , , D.M and Andrianova، نويسنده , , M.E and Sulianov، نويسنده , , S.N، نويسنده ,
Abstract :
A parallax-free X-ray area detector with a spherical Be-window has been tested in a diffractometer. The detector provides a DQE of ∼80% (8–17 keV), a parallax-free solid angle of 50°, and accuracies of 1% for the intensity and of 0.025° for the angle.