Author/Authors :
Oberstedt، نويسنده , , A and Davidsson، نويسنده , , P and Delsing، نويسنده , , P and Jonson، نويسنده , , Ji?? Kroc، نويسنده , , R and Nyman، نويسنده , , G، نويسنده ,
Abstract :
Integrated bolometers of a volume of about 0.05 mm3 are fabricated by the technique of silicon micromachining. As thermistor serves a heavily phosphorus doped area implanted in the absorber surface. The optimum thermistor doping was determined with respect to a high resistance gradient, but at the same time with resistance values low enough for providing signals which are reasonably high. Measurements of low-energetic γ-rays and conversion electrons indicated the high sensitivity of our detectors, however unfortunately for picking up noise, too. The steps to improve the current energy resolution are reported.