Title of article :
Low energy background in mercuric iodide X-ray spectrometers
Author/Authors :
Bao، نويسنده , , X.J and Natarajan، نويسنده , , M and Henderson، نويسنده , , J، نويسنده ,
Pages :
4
From page :
58
To page :
61
Abstract :
The origins of the continuous background (window effect or dead layer) in mercuric iodide X-ray spectrometers are investigated. It is found that photo-electron escape and carrier diffusion are the dominant mechanisms of incomplete charge collection in the energy range of interest (from 3–60 keV). X-ray spectra measurements, computer calculation and photo-response measurements are presented in support of the proposed model. Many observations of detector behavior made in the manufacturing and application of mercuric iodide X-ray detectors can be explained by this model.
Journal title :
Astroparticle Physics
Record number :
1999181
Link To Document :
بازگشت