• Title of article

    Electrode configuration and energy resolution in gamma-ray detectors

  • Author/Authors

    Luke، نويسنده , , P.N، نويسنده ,

  • Pages
    6
  • From page
    232
  • To page
    237
  • Abstract
    Materials used for room-temperature gamma-ray detectors exhibit substantial charge-collection deficiencies. As a result, the geometrical arrangement of charge-sensing electrodes on a detector can have a very large influence on its spectral performance. A discussion of the effects of electrode geometry on charge signal generation and energy resolution is presented. Various electrode configurations including spherical, coaxial, pixel, strip and coplanar-grid are examined in theoretical terms to evaluate their effectiveness and limitations in enhancing the performance of detectors with charge trapping.
  • Journal title
    Astroparticle Physics
  • Record number

    1999269