• Title of article

    X-ray imaging performances of a 2D hybrid cadmium telluride structure

  • Author/Authors

    Glasser، نويسنده , , F and Campagnolo، نويسنده , , R and Mathy، نويسنده , , F and Louis، نويسنده , , N and Mermet، نويسنده , , R، نويسنده ,

  • Pages
    4
  • From page
    252
  • To page
    255
  • Abstract
    A 2D imaging structure, using chlorine doped cadmium telluride connected to a read-out circuit by indium bumps, was characterized with a 70 kV dental X-ray tube. The imaging capabilities of such a structure with an experimental device of 100 μm pitch are presented. The theoretical resolution limit, due to K escape photons and photoelectrons, was computed by a Monte Carlo code simulation for 100 and 50 μm pixels. Experimental measurements of dynamic range, linearity, line spread function, signal to noise ratio and modulation transfer function are realized.
  • Journal title
    Astroparticle Physics
  • Record number

    1999278